NO2 Sensor
MiCS-2710 NO2 Sensor
This datasheet describes the use of the MiCS-2710. The package and the mode of operation illustrat...
Description
MiCS-2710 NO2 Sensor
This datasheet describes the use of the MiCS-2710. The package and the mode of operation illustrated in this document target the detection of nitrogen dioxide (NO2).
FEATURES
Low heater current
Wide detection range
High sensitivity
Fast thermal response
Miniature dimensions
High resistance to shocks and vibrations
IMPORTANT PRECAUTIONS
Read the following instructions carefully before using the MiCS-2710 sensor described in this document to avoid erroneous readings and to prevent the device from permanent damage.
The sensor must not be wave soldered without protection, or exposed to high concentrations of organic solvents, ammonia, or silicone vapours, to avoid poisoning the sensitive layer.
Heating powers above the maximum rating of 120 mW can destroy the sensor due to overheating.
This sensor is to be placed in a filtered package that protects it against any water or dust projection.
For any additional questions, email enquiries@e2v.com or telephone +44 (0)1245 493493.
OPERATING MODE
The recommended mode of operation is a constant power mode. A heater power of PH = 43 mW is applied. This causes the temperature of the sensing resistor (RS) to reach about 220 °C.
Detection of the pollution gases is achieved by measuring the sensing resistor RS during operation.
SENSOR RESPONSE
The sensor response to NO2 in air is represented in Fig. 1.
1000
100
10
0.1 1 NO2 [ppm]
10
The sensor resistance RS is normalised to the resistance under a...