AC-DC / 10W - 85W Medical
Embedded Power for Business-Critical Continuity
LPS50-M Series
60 Watts
Total Power: Input Voltage: # of Outputs:...
Description
Embedded Power for Business-Critical Continuity
LPS50-M Series
60 Watts
Total Power: Input Voltage: # of Outputs: 55 - 60 Watts 90 - 264 VAC 127 - 300 VDC Single
Rev. 1.08.09_149 LPS50-M Series of 3
Special Features
Medical Safety Approvals Universal input Less than 1U high 2” x 4” footprint Remote sense Overload and short circuit protection Adjustable output voltage High efficiency High MTBF Built in EMI filter (CISPR 22 Class B) LED power good indicator Input power < 74 watts Complies with EN61000 -3-2 UL Class I approved LPX50 enclosure kit available Dual AC fuses
Electrical Specifications
Input
Input range: Frequency: Inrush current: Input power: Efficiency: EMI/RFI: Safety ground leakage current: 90 - 264 VAC (wide range) 127-300 Vdc 47-440 Hz < 60 A peak @ 230 VAC, cold start @ 25°C < 74 Watts 80% - 85% typical at full load FCC Class B conducted; CISPR 22 Class B conducted; EN55022 Class B conducted, VDE0878PT3 Class B conducted 275uA @ 50/60 Hz; 264 VAC input
Output
Maximum power: Adjustment range: Hold-up time: Overload protection: 60W for convection (LPS52, 55 W) ±20% minimum 10/20 ms 115/230 VAC input line Short circuit protection on all outputs. Case overload protected @ 110-160% of normal rating
Overvoltage protection: 30-50% above nominal output
Safety
UL: CSA: VDE: CE: UL 60601-1 CSA-C22.2 No. 601.1-M90 EN60601-1 LVD & EMC
Remote sense:
Compensatesd for 0.5 V lead drop max. Will operate without remo...