AC-DC / 10W - 85W Medical
Embedded Power for Business-Critical Continuity
LPS40-M Series
40 Watts Medical
Total Power: Input Voltage: # of ...
Description
Embedded Power for Business-Critical Continuity
LPS40-M Series
40 Watts Medical
Total Power: Input Voltage: # of Outputs: 40-55 Watts 85-264 VAC 120-300 VDC Single
Rev. 9.8.09_28 LPS40-M Series of 3
Electrical Specifications
Special Features
Input
Input range: Frequency: Inrush current: Input current: Efficiency: EMI filter: Safety ground leakage current: 85 - 264 VAC; 120 - 300 VDC 47 - 440 Hz < 18A peak @ 115 VAC; <36A peak @ 230 VAC, cold start @ 25 °C 1A max. (RMS) @ 115 VAC 70% typical at full load FCC Class B conducted; CISPR 22 Class A conducted; EN55022 Class A conducted; VDE 0878 PT3 Class A conducted < 75µA @ 50/60 Hz, 264 VAC input
Medical safety approvals Universal input 3” x 5” footprint Less than 1U high Remote sense on main output Low output ripple Adjustable 5V output Overvoltage protection Overload protection 110 KHz switching frequency LPX40 enclosure kit available
Output
Maximum power: Adjustment range: Hold-up time: Overload protection: 40 W for convection; 55 W with 30CFM forced air -5, +10% minimum 20ms @ 40 W load, 115 VAC nominal line
Safety
VDE 0750/EN60601-1 (IEC601) UL UL2601 E182560 CSA CSA 22.2 No.601.1 LR109492 CE Mark (LVD)
Short circuit protection on all outputs Case overload protected @ 110 - 145% above peak rating Overvoltage protection: 5V output; 5.7 to 6.7 VDC. Other outputs 10% to 25% above nominal output Remote sense: ...